프로젝트 설명

온도,처리신호(전압,전류) Data Logger

SmartReader Plus 7

ACR, 데이터로거, 전압,전류 데이터로거, 전압 기록계,, 전압,전류 측정, 전압,전류 기록

Product

스마트리더플러스 7은 온도1채널 및 다양한 처리신호(전압,전류)를 7채널까지 측정,기록하는 Data Logger입니다.
7채널 아날로그 신호를 변환기로부터 자체 소프트웨어의 교정식으로 계산하여 임의의 공학값으로 변환이 가능합니다.

Specifications

Input Ranges – Standard Model 0 to 2.5 volts DC (1 channel)
0 to 5.0 volts DC (2 channels)
0 to 10.0 volts DC (1 channel)
0 to 200 millivolts DC (1 channel)
0 to 25 milliamps DC (2 channels)
Input Ranges – Other Models 0 to 2.5 volts DC (all external channels)
0 to 5 volts DC (all external channels)
0 to 10 volts DC (all external channels)
0 to 25mA DC (all external channels)
Accuracy +/- 0.5% of full scale
Overload Protection Voltage channels: +/- 40 volts (reverse-polarity protected)
Current channels: +/- 70 mA (reverse-polarity protected)
Input Impedance > 1 M ohm (200 mV channels)
> 1 M ohm (0 to 2.5 V channels)
> 1 M ohm (0 to 5.0 V channels)
> 1 M ohm (0 to 10.0 V channels)
20 ohm (0 to 25 mA channels)
Transducer Restrictions It is recommended to use a single power supply or loop isolators to avoid ground loops

Common Specifications Across Models

Size 107 x 74 x 22mm (4.2 x 2.9 x 0.9in)
Weight 110 grams (3.75 ounces)
Case Material Noryl® Plastic
Operating Limits -40°C to 70°C (-40°F to 158°F) and 0 to 95% Relative Humidity (non-condensing)
Clock Accuracy +/-2 seconds per day
Battery 3.6 volt Lithium, 1 Amp-Hour
Power Consumption 5 to 10 microamps (continuous)
Battery Life 10-year warranty (under normal use) factory replaceable
Memory Size 32 K (21,500 readings)
128 K (87,000 readings)
1.5 MB (1,000,000 readings)
See specific models for availability
Sampling Method 1. Continuous (First-in, First-out); not available with sample rates faster than 8 seconds
2. Stop when full (Fill-then-stop)
Sampling Rate User selectable rates from 25 per second to one reading every 8 hours. (BP-101 battery pack or PS-201 Power Supply required for sample rates faster than 8 seconds. See Accessories section for details.)
Resolution 12 bit (1 part in 4096)
PC Requirements Windows PC with a least 256MB RAM, 250MB of hard drive space and one free serial or USB port
Software Requirements TrendReader® 2 (compatible with Windows® XP, Vista, Windows 7, Windows 8 & Windows 10)
Product Approvals Meets Part 15 for Digital Devices of the Code of Federal Regulations of the Federal Communications Commission (FCC). Meets Class A radiated and conducted emission requirements of Section 2 of the Radio Interference Regulations of Communications Canada. IEC801-2, 3, 4 and EN55011 covering ESD, RFI Immunity, EFT/ Burst and Radiated Emissions respectively for the Commission of the European Communities (CE)
Resistance to X-Rays Tested for protection against a 160kV dose @ 5mA for 30 seconds (150mA-sec) @ 38 inches F.F.D. (about 100 times that of an airport X-Ray machine). Tested for protection against Gamma Ray (equivalent to 0.137-1.38 mega volts) IR 192 – 28 curies @ 30 second exposure source to object distance 5″