프로젝트 설명

온도,압력 Data Logger

SmartReader Plus 4

ACR, 데이터로거, 압력 데이터로거, 압력 기록계, 온도,대기압,공기압 측정, 온도,압력 기록

Product

스마트리더플러스 4는 대기압,각종 제어기기용 공기압을 측정,기록하는 Data Logger,입니다. 센서는 로거의 터미널블럭에 부착합니다. 옵션으로 내부습도 센서와 외부온도,습도 센서를 사용 가능하여 압력,온도,습도를 동시에 측정,기록 가능합니다.

Specifications

Calibrated Accuracy +/- 0.5% of full scale @ 25ºC (77ºF)
Pressure Range, Resolution and Accuracy +/- 10″ H20; 0.006″; +/- 2% FS @ 25ºC (77ºF)
0 to 5 psi (30 kPa); 0.003 psi (0.02 kPa); +/- 1% FS @ 25ºC (77ºF)
0 to 30 psi (200 kPa); 0.015 psi (0.10 kPa); +/- 0.5% FS @ 25ºC (77ºF)
0 to 100 psi (700 kPa); 0.05 psi (0.34 kPa); +/- 0.5% FS @ 25ºC (77ºF)
Accessories ACR EH-020A RH/Temperature Probe
ACR ET Series Temperature Probes or any NTC thermistor probe

Common Specifications Across Models

Size 107 x 74 x 22mm (4.2 x 2.9 x 0.9in)
Weight 110 grams (3.75 ounces)
Case Material Noryl® Plastic
Operating Limits -40°C to 70°C (-40°F to 158°F) and 0 to 95% Relative Humidity (non-condensing)
Clock Accuracy +/-2 seconds per day
Battery 3.6 volt Lithium, 1 Amp-Hour
Power Consumption 5 to 10 microamps (continuous)
Battery Life 10-year warranty (under normal use) factory replaceable
Memory Size 32 K (21,500 readings)
128 K (87,000 readings)
1.5 MB (1,000,000 readings)
See specific models for availability
Sampling Method 1. Continuous (First-in, First-out); not available with sample rates faster than 8 seconds
2. Stop when full (Fill-then-stop)
Sampling Rate User selectable rates from 25 per second to one reading every 8 hours. (BP-101 battery pack or PS-201 Power Supply required for sample rates faster than 8 seconds. See Accessories section for details.)
Resolution 12 bit (1 part in 4096)
PC Requirements Windows PC with a least 256MB RAM, 250MB of hard drive space and one free serial or USB port
Software Requirements TrendReader® 2 (compatible with Windows® XP, Vista, Windows 7, Windows 8 & Windows 10)
Product Approvals Meets Part 15 for Digital Devices of the Code of Federal Regulations of the Federal Communications Commission (FCC). Meets Class A radiated and conducted emission requirements of Section 2 of the Radio Interference Regulations of Communications Canada. IEC801-2, 3, 4 and EN55011 covering ESD, RFI Immunity, EFT/ Burst and Radiated Emissions respectively for the Commission of the European Communities (CE)
Resistance to X-Rays Tested for protection against a 160kV dose @ 5mA for 30 seconds (150mA-sec) @ 38 inches F.F.D. (about 100 times that of an airport X-Ray machine). Tested for protection against Gamma Ray (equivalent to 0.137-1.38 mega volts) IR 192 – 28 curies @ 30 second exposure source to object distance 5″